Device product
Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410
Z-1332-2021
Product summary
- Firm
- Tenacore LLC
- Event
- Event 87470
- Status
- Terminated
- Classification
- Class I
- Quantity
- 1494
- Official record key
device-enforcement:Z-1332-2021
Official wording
Reason: Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Code information: Parts distributed between July 10, 2020 and January 22, 2021. Time stamp at bezel rear side shows 5 or 6 SO Numbers: 271111, 263873, 264442, 264890, 268606, 269930, 270760, 270936, 271211, 272860, 267603, 268936, 269258, 269291, 272496, 272225, 271408, 267392
Distribution pattern: US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA
Derived failure modes
-
Unknown
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.