Skip to content
Recall Observatory FDA recall evidence

Device product

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Z-2718-2024

June 02, 2024

Class II

Product summary

Firm
Treace Medical Concepts, Inc.
Event
Event 94977
Status
Ongoing
Classification
Class II
Quantity
40 kits containing 40 plates
Official record key
device-enforcement:Z-2718-2024

Official wording

Reason: Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

Code information: UDI: (01)00810111222464(17)290424(10)300542404

Distribution pattern: US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.