Device product
Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description): MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", QF09/"Fastect¿ II Drug Screen Dipstick Test, 4 Drug Detection: THC50+COC300+OPI300+BZO300", MF20/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MTD300", MF21/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: TCA1000, MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300, MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300", QF12/ "Fastect¿ II Drug Screen Dipstick Test 4 Drug Detection: THC50+COC300+OPI300+MET500", MF10/ "Fastect¿ II Drug Screen Dipstick Test THC50", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF16 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MET500", MF23 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF14/ "Fastect¿ II Drug Screen Dipstick Test AMP1000", MF11/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: COC300", MF23 / Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100"
Z-3034-2020
Product summary
- Firm
- Ameditech Inc
- Event
- Event 86335
- Status
- Terminated
- Classification
- Class II
- Quantity
- 108,000 units
- Official record key
device-enforcement:Z-3034-2020
Official wording
Reason: Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Code information: ALL Lots Within Expiry
Distribution pattern: US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Derived failure modes
-
Unknown
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.