Skip to content
Recall Observatory FDA recall evidence

Device product

Alere iScreen Multi Drug Detection (AMT Item Number/ AMT Description): X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection:AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection: AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS10BUP-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection: AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS10BUP-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS10BUP-DX/ "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS10-DX /"Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 10 Drug Detection: Amphetamine, Barbiturate, Benzodiazepine, Cocaine, Marijuana, Methadone, Methamphetamine, Morphine, Oxycodone, and Tricyclic Antidepressant", X11-IS5 M-DX /"Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 5 Drug Detection: Amphetamine, Cocaine, Marijuana, Methamphetamine, and Morphine", X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection: AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS5 M-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 5 Drug Detection: Amphetamine, Cocaine, Marijuana, Methamphetamine, and Morphine", X11-IS10BUP-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100"

Z-3030-2020

August 18, 2020

Class II

Product summary

Firm
Ameditech Inc
Event
Event 86335
Status
Terminated
Classification
Class II
Quantity
67,300 units
Official record key
device-enforcement:Z-3030-2020

Official wording

Reason: Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Code information: ALL Lots Within Expiry

Distribution pattern: US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.