Skip to content
Recall Observatory FDA recall evidence

Device product

LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are intended for liquid quality control testing performed on the LumiraDx Instrument when used with the LumiraDx SARS-CoV-2 Ag Test Strip. The Quality Controls provide users with assurance that the device is performing within specification

Z-0135-2023

September 23, 2022

Class III

Product summary

Firm
LumiraDx
Event
Event 90998
Status
Terminated
Classification
Class III
Quantity
83 kits
Official record key
device-enforcement:Z-0135-2023

Official wording

Reason: Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC

Code information: LQC Kit: 2002204017 Positive Control Vial: 1000 1710 0014 0141 Negative Control Vial: 1000 1720 0014 0142 Exp. Date: 01-Mar-2023

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC