Device product
Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Z-0348-2024
Product summary
- Event
- Event 93222
- Status
- Ongoing
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-0348-2024
Official wording
Reason: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Code information: Model Number (UDI-DI): URF-P7 (04953170403811), URF-P7R (04953170403835), URF-V2 (04953170343582), URF-V2R (04953170343612), URF-V3 (04953170403385), URF-V3R (04953170403392).
Distribution pattern: Worldwide - US Nationwide distribution.
Derived failure modes
-
Foreign material or chemical contamination
contamination that could lead