Skip to content
Recall Observatory FDA recall evidence

Device product

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Z-0741-2022

January 18, 2022

Class II

Product summary

Firm
Cytocell Ltd.
Event
Event 89602
Status
Terminated
Classification
Class II
Quantity
19 kits
Official record key
device-enforcement:Z-0741-2022

Official wording

Reason: individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Code information: Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711

Distribution pattern: US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result