Skip to content
Recall Observatory FDA recall evidence

Device product

Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.

Z-0563-2021

November 06, 2020

Class II

Product summary

Firm
Spacelabs Healthcare, Ltd.
Event
Event 86753
Status
Terminated
Classification
Class II
Quantity
600
Official record key
device-enforcement:Z-0563-2021

Official wording

Reason: During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.

Code information: Systems that have been upgraded from Version 9 or below to Version 10 or above.

Distribution pattern: Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.