Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers.

Z-0634-2021

November 06, 2020

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 86745
Status
Terminated
Classification
Class II
Quantity
699 units
Official record key
device-enforcement:Z-0634-2021

Official wording

Reason: Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Code information: CRPL3 - lot 42402301

Distribution pattern: US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.