Skip to content
Recall Observatory FDA recall evidence

Device product

Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5

Z-0514-2021

October 30, 2020

Class II

Product summary

Firm
Invacare Corporation
Event
Event 86604
Status
Terminated
Classification
Class II
Quantity
384,767 units
Official record key
device-enforcement:Z-0514-2021

Official wording

Reason: Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

Code information: Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633; IRC5PO2VAW from serial #17HF030338-18IF018522; IRC5PO2VAWL2 from serial # 18BF007341-18IF024451; IRC5PO2VAWL3 from serial # 17JF002563-19AF008559; IRC5PO2VAWL5 from serial # 18EF003466-18EF003544. DI Number: 00841447100249

Distribution pattern: Worldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.