Skip to content
Recall Observatory FDA recall evidence

Device product

Stryker iBur 3.0mm Diamond Match Head, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-107-030D

Z-2730-2024

July 24, 2024

Class II

Product summary

Firm
Stryker Corporation
Event
Event 95040
Status
Completed
Classification
Class II
Quantity
2507 units
Official record key
device-enforcement:Z-2730-2024

Official wording

Reason: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Code information: GTIN: 07613327509229 Lot Numbers: 22334017 23047017 23095017 23198017

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.