Skip to content
Recall Observatory FDA recall evidence

Device product

Maquet Getinge- (1) Material :701067370R01 BO-TOP 38803 OPEN HEART PACK W (2) Material: 701067370R02 BO-TOP 38803 OPEN HEART PACK W (3) Material: 701067370R03 BO-TOP 38803 OPEN HEART PACK W

Z-1669-2019

April 18, 2019

Class II

Product summary

Firm
Maquet Cardiovascular, LLC
Event
Event 82703
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-1669-2019

Official wording

Reason: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Code information: Batch Numbers: (1) 3000042627 (2) 3000047034 (3) 3000058143

Distribution pattern: Nationwide

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    Sterility Assurance Level (SAL) of the stopcock ports cannot be assured