Device product

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Z-2641-2020

May 23, 2020

Class II

Product summary

Firm
Shanghai United Imaging Healthcare Co., Ltd.
Event
Event 85771
Status
Ongoing
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-2641-2020

Official wording

Reason: Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Code information: Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100

Distribution pattern: US Distribution to state of: TX

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.