Device product

Dexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1)

Z-1943-2025

May 12, 2025

Class I

Product summary

Firm
Dexcom, Inc.
Event
Event 96743
Status
Ongoing
Classification
Class I
Quantity
36,824 units
Official record key
device-enforcement:Z-1943-2025

Official wording

Reason: Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Code information: All Lots/ UDI: STK-FM-001 - 00386270003232 & STK-FR-001 - 00386270003225

Distribution pattern: worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    assembly error