Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635

Z-0599-2025

October 21, 2024

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 95666
Status
Ongoing
Classification
Class II
Quantity
11330 units
Official record key
device-enforcement:Z-0599-2025

Official wording

Reason: Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.

Code information: UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.