Skip to content
Recall Observatory FDA recall evidence

Device product

Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

Z-0173-2026

September 04, 2025

Class II

Product summary

Firm
American Contract Systems Inc.
Event
Event 97659
Status
Ongoing
Classification
Class II
Quantity
32,433 total
Official record key
device-enforcement:Z-0173-2026

Official wording

Reason: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Code information: AHLD15X UDI-DI 191072224767 Lot 8188711 Bag serial no. 17555039 17555044 17555049 17609166 17609181; AKDV96B UDI-DI 191072224019 Lot 7311311 Bag serial no. 47140875 47141919; EVBH28D UDI-DI 191072239051 Lot 8424611 Bag serial no. 17754395; UPVD01K UDI-DI 191072238993 Lot 8439111 Bag serial no. 17682598 Log 8499211 Bag serial no. 17742543

Distribution pattern: US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.