Device product
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
Z-0372-2026
Product summary
- Event
- Event 97551
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 4,272,858
- Official record key
device-enforcement:Z-0372-2026
Official wording
Reason: All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Code information: Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.
Distribution pattern: Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Derived failure modes
-
Unknown
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.