Device product
Baxter TruSystem 7500 Hybrid (MC), Product Code 1773204
Z-1416-2025
Product summary
- Event
- Event 96286
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-1416-2025
Official wording
Reason: There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Code information: UID/DI 00887761974326, All serial numbers manufactured until 11/14/2024.
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Device software or design
software issue