Device product
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Z-1390-2025
Product summary
- Firm
- Covidien
- Event
- Event 96371
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 780
- Official record key
device-enforcement:Z-1390-2025
Official wording
Reason: Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
Code information: UDI-DI/GTIN: A8845212054401, 20884521205441, 10884521205444. Lot: 202405258X. Expiration: 06/01/2029
Distribution pattern: OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan
Derived failure modes
-
Unknown
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.