Skip to content
Recall Observatory FDA recall evidence

Device product

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R

Z-1390-2025

February 26, 2025

Class I

Product summary

Firm
Covidien
Event
Event 96371
Status
Ongoing
Classification
Class I
Quantity
780
Official record key
device-enforcement:Z-1390-2025

Official wording

Reason: Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.

Code information: UDI-DI/GTIN: A8845212054401, 20884521205441, 10884521205444. Lot: 202405258X. Expiration: 06/01/2029

Distribution pattern: OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0