Skip to content
Recall Observatory FDA recall evidence

Device product

Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.

Z-0331-2024

September 25, 2023

Class II

Product summary

Firm
Aizu Olympus Co., Ltd.
Event
Event 93222
Status
Ongoing
Classification
Class II
Quantity
1243 units
Official record key
device-enforcement:Z-0331-2024

Official wording

Reason: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Code information: Model Number (UDI-DI): CYF-5 (04953170292323), CYF-VH (04953170411250), CYF-VHR (04953170411298).

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    contamination that could lead