Device product

nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers

Z-2553-2023

July 12, 2023

Class II

Product summary

Firm
Landauer
Event
Event 92647
Status
Completed
Classification
Class II
Quantity
16760 units
Official record key
device-enforcement:Z-2553-2023

Official wording

Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Code information: UDI/DI 0860003399903, all batch numbers

Distribution pattern: Worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.