Device product
Foley Catheter with temperature sensor 400TM 18FR, Catalogue Number 102201101880TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Z-2767-2018
Product summary
- Event
- Event 80641
- Status
- Terminated
- Classification
- Class II
- Quantity
- 12532
- Official record key
device-enforcement:Z-2767-2018
Official wording
Reason: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Code information: S17002333, S17002452, S17002892, S17003578, S17005045, S17003306, S17005046, S17007065, S17005298, S17000279, S17000280, S17000837, S17007171, S18001346, S17000333
Distribution pattern: Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Derived failure modes
-
Unknown
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.