Device product
Stryker iBur 4.0mm Precision Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-009-040
Z-2723-2024
Product summary
- Event
- Event 95040
- Status
- Completed
- Classification
- Class II
- Quantity
- 1227 units
- Official record key
device-enforcement:Z-2723-2024
Official wording
Reason: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Code information: GTIN: 07613327501223 Lot Numbers: 22245017 22248017 22270017 22299017 22306027 23188017
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
Derived failure modes
-
Unknown
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.