Device product
Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Z-2761-2018
Product summary
- Event
- Event 80641
- Status
- Terminated
- Classification
- Class II
- Quantity
- 16133
- Official record key
device-enforcement:Z-2761-2018
Official wording
Reason: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Code information: V17018057, V17022043, V17023777, V17018058, V17012115, V17002305, V17002306, V17002308
Distribution pattern: Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Derived failure modes
-
Unknown
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.