Device product
nanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers
Z-2548-2023
Product summary
- Firm
- Landauer
- Event
- Event 92647
- Status
- Completed
- Classification
- Class II
- Quantity
- 68007 units
- Official record key
device-enforcement:Z-2548-2023
Official wording
Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Code information: UDI/DI 0860003399903, all batch numbers
Distribution pattern: Worldwide
Derived failure modes
-
Unknown
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.