Device product
nanoDot D2DNN, Model Numbers: a) 03053-1MO; b) 03053-1WK ; c) 03053-3MO; d) 03053-OTO; e) 03053-SMO; radiation monitoring dosimeter used with the microSTAR readers
Z-2541-2023
Product summary
- Firm
- Landauer
- Event
- Event 92647
- Status
- Completed
- Classification
- Class II
- Quantity
- 35681 units
- Official record key
device-enforcement:Z-2541-2023
Official wording
Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Code information: UDI/DI 0860003399903, all batch numbers
Distribution pattern: Worldwide
Derived failure modes
-
Unknown
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.