Device product
Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1. Software: TKA 2.0.it.1+TKA 1.0.1.it.1+PKA3.0.2.it. Application Part Number: 700002190743-01
Z-1913-2024
Product summary
- Event
- Event 94473
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 68 units
- Official record key
device-enforcement:Z-1913-2024
Official wording
Reason: Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment
Code information: UDI-DI: (01)07613327629118(10). Software: TKA 2.0.it.1+TKA 1.0.1.it.1+PKA3.0.2.it. Application Part Number: 700002190743-01. Revision: AA & AB.
Distribution pattern: Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, Thailand, Turkey, UAE, and UK
Derived failure modes
-
Device software or design
Software Error #3 (SE3) error