Device product
Medfusion Syringe Pump 4000 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.
Z-1134-2018
Product summary
- Event
- Event 78922
- Status
- Terminated
- Classification
- Class II
- Quantity
- 16,600 pumps total
- Official record key
device-enforcement:Z-1134-2018
Official wording
Reason: Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.
Code information: Product Codes: 4000-0101-50, 4000-0101-51, 4000-0105-50, 4000-0105-51, 4000-0150-78, 4000-0106-00, 4000-0106-01
Distribution pattern: Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.
Derived failure modes
-
Unknown
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.