Device product
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Z-0493-2026
Product summary
- Event
- Event 97819
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 28 units (OUS only)
- Official record key
device-enforcement:Z-0493-2026
Official wording
Reason: The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Code information: Siemens Material Number (SMN): 10385891; Model/Catalog: FP3L4; UDI-DI: 00630414960418; Lot number: 629;
Distribution pattern: Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.
Derived failure modes
-
Unknown
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.