Skip to content
Recall Observatory FDA recall evidence

Device product

REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.

Z-0542-2026

October 08, 2025

Class II

Product summary

Firm
Cellavision AB
Event
Event 97835
Status
Ongoing
Classification
Class II
Quantity
688 units
Official record key
device-enforcement:Z-0542-2026

Official wording

Reason: Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.

Code information: UDI: 0735004097202/ Serial Numbers 63220-70093 or CN63220-CN70093 Serial Numbers 60001-63219 or CN60001-CN63219

Distribution pattern: US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.