Device product
STA SYSTEM CONROL N + P (ref. 00678)
Z-0738-2020
Product summary
- Event
- Event 84402
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5,437
- Official record key
device-enforcement:Z-0738-2020
Official wording
Reason: A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
Code information: REF: 00678 Lot Numbers: 253145, 253416, 253655, 254040, 254206, 254401, 254573, 254760 UDIs: (01)03607450006780(11)171130(17)191130(10)253145(241)00678 (01)03607450006780(11)180131(17)200131(10)253416(241)00678 (01)03607450006780(11)180331(17)200331(10)253655(241)00678 (01)03607450006780(11)180630(17)200630(10)254040(241)00678 (01)03607450006780(11)180731(17)200731(10)254206(241)00678 (01)03607450006780(11)180930(17)200930(10)254401(241)00678 (01)03607450006780(11)181031(17)201031(10)254573(241)00678 (01)03607450006780(11)181130(17)201130(10)254760(241)00678
Distribution pattern: US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Derived failure modes
-
Unknown
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.