Device product
myQA iON; Article Number: MQ10-000;
Z-1703-2025
Product summary
- Event
- Event 96647
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 84 units (9US, 75 OUS)
- Official record key
device-enforcement:Z-1703-2025
Official wording
Reason: Wrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.
Code information: Article Number: MQ10-000; UDI-DI: EIBAMQ010000;
Distribution pattern: Worldwide - US Nationwide distribution in the states of AL, AR, FL, GA, KS, MD, OK and the countries of Algeria, Austria, Belgium, Czech Republic, Denmark, France, Germany, Ireland, Italy, Morocco, Norway, Poland, Slovakia, Spain, Israel, Kazakhstan, Kosovo, Libya, South Africa, Turkey, United Kingdom, Switzerland, Argentina, Brazil, Mexico, APAC: Indonesia, Japan, Peoples Republic of China, Philippines, South Korea, Thailand.
Derived failure modes
-
Unknown
Wrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.