Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Z-0016-2026

August 29, 2025

Class II

Product summary

Firm
Getinge Disinfection Ab
Event
Event 97535
Status
Ongoing
Classification
Class II
Quantity
902 units
Official record key
device-enforcement:Z-0016-2026

Official wording

Reason: Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Code information: Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.

Distribution pattern: Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.