Device product

Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500

Z-2477-2026

May 01, 2026

Class II

Product summary

Firm
GE Medical Systems, LLC
Event
Event 98986
Status
Ongoing
Classification
Class II
Quantity
265 units
Official record key
device-enforcement:Z-2477-2026

Official wording

Reason: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Code information: Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.