Device product

Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;

Z-1341-2026

January 13, 2026

Class II

Product summary

Firm
Laerdal Medical Corporation
Event
Event 98039
Status
Ongoing
Classification
Class II
Quantity
1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
Official record key
device-enforcement:Z-1341-2026

Official wording

Reason: units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Code information: 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0