Skip to content
Recall Observatory FDA recall evidence

Device product

Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005

Z-2124-2026

January 07, 2026

Class II

Product summary

Firm
Medline Industries, LP
Event
Event 98642
Status
Ongoing
Classification
Class II
Quantity
240 units
Official record key
device-enforcement:Z-2124-2026

Official wording

Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code information: 1) DYNJ35363A, UDI-DI: 10884389516461(each), 40884389516462(case), Lot Number: 21BBP020; 2) SPTPCD10005, UDI-DI: 10080196228652(each), 40080196228653(case), Lot Number: 24JBB041

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.