eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151

Z-1733-2026

March 02, 2026

Class II

Product summary

Firm
Civco Medical Instruments Co. Inc.
Event
Event 98513
Status
Ongoing
Classification
Class II
Quantity
18 units
Official record key
device-enforcement:Z-1733-2026

Official wording

Reason: There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Code information: UDI-DI: 00841436111058; Lot Number: A228675, A228676, A233854, A268115, A268237, A272509, A277538, A280341

Distribution pattern: US States: OH, PA, WA. China.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.