Device product
HemosIL LMW Heparin Controls; Part Number: 0020300200;
Z-2323-2025
Product summary
- Event
- Event 97093
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 41,824 units
- Official record key
device-enforcement:Z-2323-2025
Official wording
Reason: Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Code information: Part Number: 0020300200; UDI-DI: 08426950472490; Lot Numbers: N0330228, N0632773, N0935135, N1036592, N0148986, N0441476, N0643673, N0845161, N1147397;
Distribution pattern: Worldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Macao, Malaysia, Mexico, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Vietnam.
Derived failure modes
-
Unknown
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.