Device product

Apo B Reagent, REF: OSR6143

Z-1815-2025

April 23, 2025

Class III

Product summary

Firm
Beckman Coulter Ireland, Inc.
Event
Event 96746
Status
Ongoing
Classification
Class III
Quantity
2747 units
Official record key
device-enforcement:Z-1815-2025

Official wording

Reason: The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

Code information: Lot # 2631/UDI: 15099590010409

Distribution pattern: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.