Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Z-1663-2025

March 27, 2025

Class II

Product summary

Firm
IHB OPERATIONS B.V.
Event
Event 96581
Status
Ongoing
Classification
Class II
Quantity
112 units
Official record key
device-enforcement:Z-1663-2025

Official wording

Reason: The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Code information: UDI/DI 10841736116071, All Serial Numbers

Distribution pattern: US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.