Skip to content
Recall Observatory FDA recall evidence

Device product

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Z-2141-2024

May 16, 2024

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 94676
Status
Ongoing
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-2141-2024

Official wording

Reason: Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

Code information: UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087

Distribution pattern: US Nationwide distribution in the states of AL, IN, AK, WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0