Skip to content
Recall Observatory FDA recall evidence

Device product

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Z-1303-2024

February 08, 2024

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 94012
Status
Ongoing
Classification
Class II
Quantity
64,009 kits
Official record key
device-enforcement:Z-1303-2024

Official wording

Reason: Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

Code information: Lot numbers: 338555, exp. 31 JUL 2024; 338556, exp. 31 AUG 2024; 338734, exp. 30 SEP 2024; and 338958, exp. 31 OCT 2024; UDI-DI 15099590225742.

Distribution pattern: Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.