Device product

VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of vitamin B12 - Product Code: 6800872

Z-2019-2024

May 08, 2024

Class III

Product summary

Firm
Ortho-Clinical Diagnostics, Inc.
Event
Event 94650
Status
Ongoing
Classification
Class III
Quantity
76 units
Official record key
device-enforcement:Z-2019-2024

Official wording

Reason: The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.

Code information: UDI-DI: 10758750000937 Lot Number: 1200 Exp. Date: 30-Sep-2024

Distribution pattern: Worldwide - US Nationwide and the country of Columbia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.