Skip to content
Recall Observatory FDA recall evidence

Device product

Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Z-0714-2024

December 05, 2023

Class II

Product summary

Firm
Ambu Inc.
Event
Event 93599
Status
Ongoing
Classification
Class II
Quantity
11845 distributed US
Official record key
device-enforcement:Z-0714-2024

Official wording

Reason: Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

Code information: Catalog No. 405011000 - UDI-DI 5707480145089 Catalog No. 405011000US2 - UDI-DI 5707480156757 All products distributed since launch (May 2020)

Distribution pattern: Domestic distribution nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.