Skip to content
Recall Observatory FDA recall evidence

Device product

Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

Z-0290-2024

October 12, 2023

Class II

Product summary

Firm
Maquet Medical Systems USA
Event
Event 93013
Status
Ongoing
Classification
Class II
Quantity
1022 units
Official record key
device-enforcement:Z-0290-2024

Official wording

Reason: It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

Code information: UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868

Distribution pattern: US Nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.