enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module and TCAutomation Software Version 3.6.1 and Below; IVD. --- Thermo Scientific Centrifuge Module has Product Code 952040-EG 6844097. The VITROS 5,1 FS Chemistry System with enGen Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).

Z-0300-2017

September 07, 2016

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 75167
Status
Terminated
Classification
Class II
Quantity
60 (US: 24; Foreign: 36)
Official record key
device-enforcement:Z-0300-2017

Official wording

Reason: Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the laboratory prior to the identification of the Cross Check error on the enGen Laboratory Automation System. Assay results associated with the wrong patient and reported out of the laboratory could lead to inappropriate intervention with the potential for serious injury to the patient.

Code information: Software Versions 3.6.1 and below; Affected US Serial Numbers: 5GV6W2J,GBX395J, J2QKG25J, 2Z3YV1J, 3RHC23J, 4HX7S4J, 54FC01J, 5J4B45J, 736BM4J, 7KSC15J, J86PWC5J, 953209, 953221, 953266, 953270, B3ZHP2J, CXL941J, JFBX395J, FCM073J, FXCBG3J, JG7BFB5J, GWXQ52J, HJ5K84J, HPKG25J -- Affected Foreign Serial Numbers: DWXO52J, GWJZY4J, J16XQH4J, BGV6W2J, 17QP94J, 4SHPF4J, 1K5K84J, 37QP34J, H5PWC5J, J2V6LW4J, F6MXR3J, JT6LW4J, N0551107, F1150698, 46XQH4J, 2S5OP0J, CRKM82, J41BCY4J, J56T205J, BGV6W2J, 7YCF43J, 96XQHAJ, C1JNZ5J, C3JTP4J, D3JTP4J, 86T205J, 18BFB5J, 96PWC5J, 1LSSF5J, 4drj44j, 8W7WG3J, J953222, J953244, JCT20Q3J, HWJZY4J, 38769917923.

Distribution pattern: Worldwide Distribution-(Nationwide) and foreign distribution to the following countries: Belgium, Bermuda, Canada, Chile, Denmark, France, Italy, Mexico, Norway, Portugal, Sweden, Taiwan and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the laboratory prior to the identification of the Cross Check error on the enGen Laboratory Automation System. Assay results associated with the wrong patient and reported out of the laboratory could lead to inappropriate intervention with the potential for serious injury to the patient.