Skip to content
Recall Observatory FDA recall evidence

Device product

Kimberly-Clark MIC PEG Kit - Pull Type The Kimberly-Clark MIC PEG Kits - Pull type are Percutaneous Endoscopic gastrostomy (PEG) tubes intended for a non-surgical "pull" endoscopic placement technique. The PEG tube is held in place during use with an external retention disk.

Z-2416-2012

June 22, 2010

Class II

Product summary

Firm
Kimberly-Clark Corporation
Event
Event 63001
Status
Terminated
Classification
Class II
Quantity
766 units
Official record key
device-enforcement:Z-2416-2012

Official wording

Reason: The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

Code information: Lots AW916002A, AW916102D

Distribution pattern: Nationwide Distribution including AL, AK, AZ, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, ME, MA, MI, MN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA, WA, and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0