ONCOR Impression Plus, Digital Linear Accelerator, Model No. 05857912 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1401-2018

February 14, 2018

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 79766
Status
Terminated
Classification
Class II
Quantity
15
Official record key
device-enforcement:Z-1401-2018

Official wording

Reason: Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information: Serial no: 3855 4019 4067 3953 5148 3793 5098 5222 3801 5388 3948 3755 5033 4015 5095

Distribution pattern: Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.