Device product

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Z-2255-2016

June 22, 2016

Class II

Product summary

Firm
Olympus Scientific Solutions Americas
Event
Event 74593
Status
Ongoing
Classification
Class II
Quantity
2, 210 units distributed in the USA
Official record key
device-enforcement:Z-2255-2016

Official wording

Reason: Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

Code information: All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0