Device product
Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
Z-2949-2020
Product summary
- Event
- Event 85822
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 67 units
- Official record key
device-enforcement:Z-2949-2020
Official wording
Reason: One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Code information: All lots.
Distribution pattern: U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.
Derived failure modes
-
Unknown
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.